How to Calibrate an O2 Gas Sensor?

How do I calibrate the Go Direct O2 Gas?
The sensor is factory calibrated. For many experiments, it will not be necessary to calibrate. However, for more accurate measurements, the sensor can be calibrated using a one-point calibration using a known concentration of atmospheric oxygen.
To calibrate the sensor with Graphical Analysis or Graphical Analysis Pro:
- Connect the sensor to the application.
- Once connected, select O2 Gas button in the bottom right hand corner and select Calibrate.
- Enter your known oxygen concentration value based on humidity or enter 20.9 and click Keep.
- Then click Apply.
To calibrate the sensor with LabQuest App:
- Connect the sensor to the device.
- Once connected, tap on the live O2 gas meter, then select Calibrate
- Tap on Calibrate Now and enter your known oxygen concentration value based on humidity or enter 20.9
- Tap Apply.
The calibration will be automatically saved to the sensor.
How do I calibrate the O2 Gas Sensor?
A brand new O2 Gas Sensor does not need to be calibrated, but the sensor will drift over time and will need to be calibrated.
To calibrate the sensor, follow the normal 2-point calibration procedure.
- For the first point, push and hold the calibrate button on the sensor using a paper clip or LabQuest stylus.
- Enter a value of 0 for this reading.
- Release the button and take a second reading in air.
- Enter a value of 20.9% oxygen or a corrected value from the table in the sensor booklet.
Since the timing of when you press and release the calibrate button on the sensor is important, the video should be every helpful.