How to care for your pH Sensors

How to care for your pH Sensors
Using a pH sensor is an excellent resource for enabling your students to delve into the scientific principle known as pH.
Nethertheless, pH sensors demand a little extra care to prolong their lifespan. We will provide you with methods on how to store your pH sensors and guide you on how to calibrate them.
The Best Way to Store pH Sensors
The most important part of storing your pH sensors is to use the storage solution which is provided and make sure that the solution is replaced around twice a year. The storage bottle should always be kept three quarters full of the solution to keep the sensor’s bulb submerged even when stored horizontally. If you are going to not use your sensors for an extended period of time, it is advisable to replace the storage solution beforehand to prevent mould growth.
Where to get pH or ORP Storage Solution?
We can supply a pH/ORP Storage Solution which has a mould inhibitor added. Each bottle contains 500 mL of pH/ORP Storage Solution.
If you wish to prepare your own storage solution, you can do so by adding 10g of solid potassium chloride (KCl) to 100 mL of a buffer pH 4 solution. It is also a good idea to add some buffer preservative (add 3 drops preservative/100 mLs solution). We currently ship our electrodes in a buffer storage solution that already has Buffer Preservative added to it.
If you do not have pH 4 buffer solution, we do offer pH Buffer Capsule Kit. If you don’t have this available, you can prepare your own pH 4 buffer using the following recipe: Add 2.0 mL of 0.1 M HCl to 1000 mL of 0.1 M potassium hydrogen phthalate.
What do I do about mould growth in my bottle of pH electrode storage solution?
Mould growth normally appears as small blackish or grayish specks in the electrode storage solution, and it can also stick to the walls of the bottle and the sensor tip. It does not affect the accuracy or the response of the sensor, but it should be cleaned. Rinse the bottle and its cap with warm soapy dish detergent. If it still persists, it may be followed by a brief (2-3 minutes) soak of 1-2 drops of grocery store bleach and a few mL of distilled water in the bottle to kill any remaining bugs. If you have persistent mould growth in the storage solution bottles, the pH storage bottle and lid can be cleaned with straight household bleach as long as it gets a thorough rinsing afterwards to remove any traces of remaining bleach.
To prevent mould growth, it is a good idea to replace the pH electrode storage solution with fresh solution once during the school year, and one more time before storing the sensors for the summer. The recipe for storage solution can be found above.
How do I calibrate my pH sensor?
A calibration equation is stored on each Vernier pH sensor before it is shipped. For the most accurate measurements with your sensor, perform a two- or three-point calibration with two or three different buffer solutions. For best results, the calibration points should be widely separated and bracket the range you anticipate in your experiment. For most chemistry experiments, the buffer solutions typically have a pH of 4, 7, and/or 10.
Before doing a two- or three-point calibration, you must prepare your sensor by removing the storage bottle and rinsing the tip with deionized water. Don’t wipe the sensing tip. Rather, blot it with a lint-free tissue to remove excess moisture—and don’t wipe the sensing glass.
If the glass bulb of my pH sensor is broken, can it be repaired?
Unfortunately, no. However, some Vernier pH sensors can be purchased in two parts, so you can replace just the electrode. For instance, the Go Direct® pH BNC Electrode is the replacement electrode for the Go Direct pH Sensor.
Here are some tips to help keep the glass bulb of your pH sensor intact:
- Avoid hitting the edge of the beaker or other vessel with the tip of the electrode.
- Refrain from using the electrode to stir a solution. If you must do so, avoid touching the bottom and sides of the vessel.
- If you are using a stir bar, do not lower the tip of the electrode so far that the stir bar strikes it.
- Never let the glass bulb touch anything other than liquid. For example, if you’re measuring the pH of a stream, don’t let the tip of the sensor touch rocks, sand, or gravel.
I found an old pH sensor in a supply closet. How do I know if it works?
Start with a calibration check (see above). If that fails, follow these three steps to rejuvenate, or shock, your older pH sensor:
- Soak the sensor in 0.5 M hydrochloric acid overnight.
- Rinse the sensor and soak it in fresh storage solution for one hour.
- Calibrate the sensor and store the calibration.
If your pH sensor doesn’t work after you shock it, you may need to replace it and dispose of the old one. But here’s the good news: If your pH sensor comes in two parts, you have to replace only the electrode portion. Also, with good care and feeding, a pH sensor can last about five to seven years.
And if you have additional questions about pH sensor care—or Vernier pH sensors—we’re happy to help! Feel free to contact us at support@inds.co.uk or 01530 832 500.