Measuring Ion Concentration by Colorimetry- Beer’s Law

CHEMISTRY

Measuring Ion Concentration by Colorimetry- Beer’s Law


This Experiment is based on the Chemistry with Vernier Lab book  Experiment 11.The primary objective of this experiment is to determine the concentration of an unknown copper sulfate solution. To accomplish this, you will use a Colorimeter or a Spectrometer to pass light through the solution, striking a detector on the opposite side. The wavelength of light used should be one that is absorbed by the solution. The CuSO4 solution used in this experiment has a deep green color, so Colorimeter users will be instructed to use the red LED. Spectrometer users will determine an appropriate wavelength based on the absorbance spectrum of the solution. The light striking the detector is reported as absorbance or percent transmittance. A higher concentration of the colored solution absorbs more light (and transmits less) than a solution of lower concentration.

You are to prepare five copper sulfate solutions of known concentration (standard solutions). Each is transferred to a small, rectangular cuvette that is placed into the Colorimeter or Spectrometer. The amount of light that penetrates the solution and strikes the detector is used to compute the absorbance of each solution. When a graph of absorbance vs. concentration is plotted for the standard solutions, a direct relationship should result. The direct relationship between absorbance and concentration for a solution is known as Beer’s law.

The concentration of an unknown CuSO4 solution is then determined by measuring its absorbance. By locating the absorbance of the unknown on the vertical axis of the graph, the corresponding concentration can be found on the horizontal axis. The concentration of the unknown can also be found using the slope of the Beer’s law curve.

Objectives

In this experiment, you will

  • Prepare CuSO4 standard solutions.
  • Measure the absorbance value of each standard solution.
  • Find the relationship between absorbance and concentration of a solution.
  • Use the results of this experiment to determine the unknown concentration of another CuSO4 solution.

For preparation of solutions

  • Volumetric flasks and test tubes
  • Salt (e.g. CuSO4.5H2O)
  • Pipette
  • Distilled water
  • Balance

For measurement and analysis

Solution Preparation.

Make up 5 solutions over a range of concentrations e.g. 2 M to 0.2 M. label 5 cuvettes with these concentrations and fill them ¾ full with the corresponding solutions. A set of cuvettes should also be prepared for testing against these standards, and labelled A, B and C, with solutions of concentrations unknown by the student and within the range of the reference solutions e.g. 2 M to 0.2 M. Ensure that the clear sides of the cuvettes are clean and avoid handling the cuvettes by the clear sides during the process—clean with a microfibre or lens cloth if necessary.

Method

On the Vernier Go Direct Colorimeter, press the power button once. The LED will blink red. Either connect the Vernier Go Direct Colorimeter to the computer via a USB cable or ensure Bluetooth is enabled on the computer and launch the Graphical Analysis app. From the file icon in the top left of the app window, select New Experiment. A Bluetooth connection will now ask you to choose the Vernier Go Direct Colorimeter from a list and select done (note the serial number if you have more than one listed); the LED on your Vernier Go Direct Colorimeter will blink green, indicating that you are ready to record data. A USB connection will take you straight to the graph view and a solid orange LED indicates that the USB connection is made.

Select Mode from the bottom left of the graph screen. Change from Time Based to Event Based in the top of dialogue box. In the Event Name field type Volume, and in the Units field cm3. You must now calibrate the the colorimeter using a cuvette filled approximately ¾ full with distilled water. Slide the colorimeter cover open, place the cuvette in the chamber, ensuring that the clear side aligns with the embossed arrow to the side of the chamber, and close the cover. Use the arrow keys on the colorimeter to select the red wavelength (635 nm)—red, because the ions we are measuring are blue so more red light will be absorbed. Then press CAL on the colorimeter, observe the red LED flashing and verify that the Absorbance value shown at the bottom of the iPad screen is 0.00.

You are now ready to collect data. First select Collect from the top of the graph screen. A graph and table are now displayed, along with Keep and Stop buttons. Place the cuvette containing the 2.0 M solution in the chamber and close the cover, as before. Click on Keep and record the concentration. Click Done and the concentration and absorbance are recorded in the table and plotted on the adjacent graph. Repeat the process for each of the known concentrations of copper sulphate solution so that five points are shown on the graph.

Analysis

Tap on the windowed button, top right; select 1 Graph. Click on the Graph Tools button, bottom left, switch on interpolate, select Apply Curve Fit and choose linear. Now take the cuvette containing sample A, the solution of unknown concentration, and place in the colorimeter as before. The Absorbance of this sample is displayed beneath the graph. Tap on the graph once and a vertical reference line will appear, showing the absorbance value next to the line and the concentration value about ¾ of the way up the screen. Click and drag this vertical line to match the current value displayed for absorbance of sample A, and read off the concentration (if the two boxes are overlying, such that the values cannot be read, open the Graph Options, in Graph Tools, and select Edit Graph Options. Here you may set the scale of the y-axis such that the two readings are separated). Repeat for the other unknown samples.