Navigation

Calibrating Climatic Chambers

Almemo Measuring System for Calibrating Climatic Chambers

Climatic chamber calibration system comprising of an Almemo® 710 instrument, one Pt100 Psychrometer and eight Pt100 temperature sensors are used to acquire all relevant measurable variables prevalent in the climatic chamber. The relative atmospheric humidity at the nine points in the climatic chamber is calculated in the ALMEMO® 710 itself. Climatic chambers can thus be calibrated in full and on site quickly and easily. Humidity is calculated in the ALMEMO® 710 on the basis of formulae as per Dr. Sonntag and the enhancement factor as per W. Bögel (correction factor Fw(t,p)) for real mixed gas systems). This substantially widens the measuring range and improves the accuracy of humidity variable calculations.

All values, both measured and calculated, are shown in a clear and easy-to-understand way on the ALMEMO® 710‘s large touch display. The ALMEMO® 710 also operates as a data logger. Measuring series can be saved either to the internal memory (capacity for over 400,000 measured values) or via the ALMEMO® memory connector to an SD card (capacity for several millions of measured values). WinControl can be used to display and document values e.g. as a line graphic – either online those measured values actually being acquired during a measuring operation or offline after a measuring operation those measured values previously saved. It also provides various evaluation and statistical functions

Precision Measuring instrument ALMEMO® 710

10 inputs for any ALMEMO® sensors, atmospheric pressure sensor integrated in the measuring instrument.

Precision measuring instrument ALMEMO® 710
including USB cable, mains unit, instrument case, and confi guration software ALMEMO® Control
MA710
DAkkS calibration certificate for atm. pressure sensor five points in range 700 to 1100 mbarKD9213D

Pt100 Psychrometer

Operative range 0 (not ice) to 90 °C, 10 to 100 % RH
The psychrometer is positioned at the center of the useful volume. From the measured values – dry temperature (t) and wet temperature (tw) – and atmospheric pressure (p) (atmospheric pressure sensor integrated in the ALMEMO® 710) we can calculate the relative humidity (Uw) at the center and the dewpoint (td).

Pt100 psychrometer FPAD363 including mains unit, water bottle, one pair of wicksFPAD363
Programming for digital psychrometer: Dew point tdOA9000PRTD
DAkks calibration certifi cate for atmospheric humidity
Two climate points at 25°C, 30%RH and 25°C, 75%RH (other points available on request)
KH9146D
DAkkS calibration certificate for air pressure sensor, 5 points in the range 700-1100 mbKD9213D
Case for psychrometer and accessoriesZB2490TK2

Pt100 temperature sensors

For operation in the climatic chamber stainless steel protective tube with PFA cable. Operative range -100 to +250 °C, Protective class IP68 The eight temperature sensors are positioned at the corners of the cuboid spanning the useful volume. From the eight measured values for temperature (t) and the humidity variables from the psychrometer we can calculate the relative humidity values (Uw) at the corners of the cuboid

Eight Pt100 temperature sensors, diameter 4mm, for operation in the climatic chamber,
IP68, Cable length = 5 meters
8 x FPA40ST0050S01KL0050
DAkks calibration certificate for temperature, three points at 0, 50, 100 °C
(other points available on request)
for 1st sensor1 x KT9021D
for 2nd to 8th sensor7 x KT9021D2
Multi-point adjustment for eight sensors
(in certificate, sensor deviation virtually reduced to zero)
8 x KA9001DW
Programming for eight Pt100 temperature sensors for calculating humidity using ALMEMO® 710 including labeling of the sensor connectorOA9000PRKS

Determination of the influence of radiation

Two temperature sensors with different surfaces to determine the radiation effects on air temperature measurement operations. (different emissivity e.g. stainless steel and PTFE)

Cover for Pt100 temperature sensor, diameter 4mm, PTFE, large emissivityZT9000TS41

With an ALMEMO® double sensor, the 2 temperatures can be measured simultaneously (additionally) with the 8 temperatures of the corners.

Digital Pt100 temperature sensors, diameter 4 mm each, for use in climatic cabinet, IP68, cable length 5 m each, mounted on 1 ALMEMO® D7 double connectorFPD740ST0050S21KL0050
DAkkS calibration certificate for temperature, 3 points at 0 / 50 / 100 °C (other points on request)
for 1st sensorKT9021D
for 2nd sensorKT9021D2
Multi-point adjustment for 2 probes (with certificate probe deviation towards zero)2 x KT9001DW