| Type of Sensor | Model | Meas. Range | Dim. | Resol. | Linearisation Accuracy | Connector Progr. |
| Resistance-based temperature sensors |
| Pt100/1000-1 4-conductor | FP Axxx | -200.0 … +850.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9030-FS1 / 4 |
| Pt100/1000-2 4-conductor | FP Axxx | -200.0 … +400.0* | °C | 0.01 K | ±0.05 K | ZA 9030-FS2 / 5 |
| Pt100-3 4-conductor | FP Axxx | 8.000 … +65.00* | °C | 0.001 K | ±0.002 K | ZA 9030-FS7 |
| Ni100/1000 4-conductor | | -60.00 …+240.00 | °C | 0.1 K | ±0.05 K | ZA 9030-FS3 / 6 |
| Ntc type N | FP Axxx | -50.00 …+125.00 | °C | 0.1 K | ±0.05 K | ZA 9040-FS |
| Thermocouples |
| NiCr-Ni (K) | FT Axxx | -200.0 …+1370.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9020-FS |
| NiCroSil-Nisil (N) | -200.0 …+1300.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9021-FSN |
| Fe-CuNi (L) | -200.0 …+900.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9021-FSL |
| Fe-CuNi (J) | -200.0 …+1000.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9021-FSJ |
| Cu-CuNi (U) | -200.0 …+600.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9000-FSU |
| Cu-CuNi (T) | -200.0 …+400.0 | °C | 0.1 K | ±0.05 K ±0.05 % of meas.v. | ZA 9021-FST |
| PtRh10-Pt (S) | 0.0 …+1760.0 | °C | 0.1 K | ±0.3 K | ZA 9000-FSS |
| PtRh13-Pt (R) | 0.0 …+1760.0 | °C | 0.1 K | ±0.3 K | ZA 9000-FSR |
| PtRh30-PtRh6 (B) | +400.0 …+1800.0 | °C | 0.1 K | ±0.3 K | ZA 9000-FSB |
| AuFe-Cr | -270.0 …+60.0 | °C | 0.1 K | ±0.1 K | ZA 9000-FSA |
| Electrical and Digital Signals |
| Millivolt DC | -10.0 …+55.0 | mV | 1µV | - | ZA 9000-FS0 |
| Millivolt 1 DC | -26.0 …+26.0 | mV | 1µV | - | ZA 9000-FS1 |
| Millivolt 2 DC | -260.0 …+260.0 | mV | 0.01 mV | - | ZA 9000-FS2 |
| Volt DC | -2.6 …+2.6* | V | 0.1 mV | - | ZA 9000-FS3 |
| Volt DC | -26.0 …+26.0 | V | 1 mV | - | ZA 9602-FS |
| For measuring bridges 5V (example) | -26.0 …+26.0 | mV | 1 µV | - | ZA 9650-FS1V |
| For potentiometer, supply 2.5V | -2.6 …+2.6* | mV | 0.1 mV | - | ZA 9025-FS3 |
| Volt AC (50Hz...2kHz) (example) | 0 …+26 | V | 0.1 V | - | ZA 9603-AK3 |
| Volt AC (11Hz...250Hz) (example) | 0 …+400 | V | 1 V | - | ZA 9903-AB5 |
| Ampere AC (11Hz...250Hz) (example) | 0 …+10.00 | A | 0.01 A | - | ZA 9904-AB2 |
| Volt DC (sampling rate 1kHz)(example) | 0 …+400 | V | 1 V | - | ZA 9900-AB5 |
| Ampere (sampling rate 1kHz)(example) | 0 …+10.00 | A | 0.01 A | - | ZA 9901-AB4 |
| Milliampere DC | -32.0 …+32.0 | mV | 1 µA | - | ZA 9601-FS1 |
| Percent(4-20mA DC) | 0.0 …100.0 | % | 0.01 % | - | ZA 9601-FS2 |
| Ohm | 0.00 …500.00* | Ω | 0.01 Ω | - | ZA 9003-FS |
| Ohm | 0.0 …5000.00* | Ω | 0.1 Ω | - | ZA 9003-FS2 |
| Frequency | 0 … 15000 | Hz | 1 Hz | - | ZA 9909-AK1U |
| Pulses/measuring cycle | 0 … 15000 | | | - | ZA 9909-AK2U |
| Digital interface | 0 … 15000 | | | - | ZA 9919-AKxx |
| Digital input | 0.00… 100.00 | % | | - | ZA 9000-ES2 |
| Capacitive Humidity Sensors |
| Relative humidity | FH A646 | 5.0 … 98.0 | %H | 0.1 % | - | |
| Relative humidity with TC | FH A646-R/C | 5.0 … 98.0 | %H | 0.1 % | ±0.5 % | |
| Dew point temperature | -25.0 … 100.0 | °C | 0.1 K | ±0.2 K | |
| Mixture ratio | 0.0 … 500.0 | g/kg | 0.1 g/kg | ±0.5 % of meas.v. | |
| Partial vapour pressure | 0.0 … 1013.2 | mbar | 0.1 mbar | ±0.1mbar ±0.1 % of meas.v. | |
| Enthalpy | 0.0 … 400.0 | kJ/kg | 0.1 kJ/kg | ±0.5 % of meas.v. | |
| Psychrometer | FN A846 | | | | | ZA 9846-AK |
| Humid temperature | 0.00 …+100.00 | °C | 0.01K | ±0.05 K | |
| Relative humidity | 0.0 … 100.0 | %H | 0.1 % | ±1.0 %H | |
| Dew point temperature | -25.0 … 100.0 | °C | 0.1 K | ±0.2 K | |
| Mixture ratio | 0.0 … 500.0 | g/kg | 0.1 g/kg | ±0.5% of meas.v | |
| Partial vapour pressure | 0.0 … 1013.2 | mbar | 0.1 mbar | ±0.1mbar ±0.1% of meas.v. | |
| Enthalpy | 0.0 … 400.0 | kJ/kg | 0.1 kJ/kg | ±0.5% of meas.v. | |
| Flow Sensor |
| Rotating vane, normal | FV A915-S120 | 0.30 … 20.00 | m/s | 0.01 m/s | ±0.1 m/s ±0.2% of meas.v. | ZA 9915-AKS1 |
| Rotating vane, normal | FV A915-S140 | 0.40 … 40.00 | m/s | 0.01 m/s | ±0.2 m/s ±0.2% of meas.v. | ZA 9915-AKS2 |
| Rotating vane, normal | FV A915-S220 | 0.50 … 20.00 | m/s | 0.01 m/s | ±0.1 m/s ±0.2% of meas.v. | ZA 9915-AKS3 |
| Rotating vane, normal | FV A915-S240 | 0.60 … 40.00 | m/s | 0.01 m/s | ±0.2 m/s ±0.2% of meas.v. | ZA 9915-AKS4 |
| Rotating vane, normal | FV A915-MA1 | 0.10 … 20.00 | m/s | 0.01 m/s | ±0.1 m/s ±0.2% of meas.v. | ZA 9915-AK5 |
| Rotating vane, normal | FV A915-WM1 | 0.00 … 5.00 | m/s | 0.01 m/s | ±0.1 m/s ±0.2% of meas.v. | ZA 9915-AK6 |
| Dyn. pressure sensor | FD A602-S1K | 0.5 … 40.0 | m/s | 0.1 m/s | ±01. m/s | |
| Dyn. pressure sensor | FD A602-S6 | 1.8 … 90.0 | m/s | 0.1 m/s | ±01. m/s | |
| Thermoanemometer | FV A935-TH4 | 0 … 2.000 | m/s | 0.001 m/s | - | |
| Thermoanemometer | FV A9355-TH3 | 0 … 20.00 | m/s | 0.01 m/s | - | |
| Thermoanemometer | FV A605-TA1 | 0.01 … 1.000 | m/s | 0.001 m/s | - | |
| Thermoanemometer | FV A605-TA5 | 0.15 … 5.00 | m/s | 0.01 m/s | - | |
| Chemical Probes |
| Conductivity | FY A641-LF | (e.g.)0.0 … 20.000 | mS | 0.001 mS | ±0.2% of meas.v. | |
| O2 dissolved, saturation | FY A640-O2 | 0 … 260 | % | 1% | - | |
| O2 dissolved, concentration | FY A640-O2 | 0.0 … 40.0 | mg/l | 0.1 mg/l | ±0.2 mg/l | |
| O2 in gases | FY 9600-O2 | 1 … 100 | % | 1% | - | |
| O3 in gases | FY 9600-O3 | 0 … 300 | ppb | 20 ppb | - | |
| CO probe | FY 9600-CO | 0 … 300 | ppm | 1 ppm | - | |
| CO2 in gases | FY A600-CO2 | (e.g.)0.000… 0.500 | % | 0.01 % | ±0.2 % of meas.v. | |
| pH-probe | FY 96PH-Ex | 0.0 … 14.00 | pH | 0.01 pH | - | ZA 9610-AKY4W |
| Redox probe | FY 96RX-Ex | 0.0 … 2600.0 | mV | 0.1 mV | - | ZA 9610-AKY5W |
| Optical Radiation (examples) |
| Lux measuring probe | FL A613-VL | 0 … 260000 | lux | 1 lux | - | |
| Lux measuring probe | FL A603-VL2 | 0.5 … 12500 | lux | 0.01 lux | - | |
| Lux measuring probe | FL A603-VL4 | 1 … 250000 | lux | 1 lux | - | |
| UV measuring probe | FL A613-UV | 0 … 87.00 | W/m2 | 0.01 W/m2 | - | |
| UVA measuring probe | FL A603-UV24 | 0.0004 … 100 | mW/cm2 | 0.1 µW/cm2 | - | |
| Radiometric meas. head | FL A603-RW4 | 0.00004 … 10 | mW/cm2 | 0.1 µW/cm2 | - | |
| Photosynthesis meas. head | FL A603-PS5 | 0.0002 … 100 | mmol/m2s | 0.1µmol/m2s | - | |
| Further transducers that can be connected (examples) |
| Heat flow plates | FQ Axxx | -260 … +260 | mV | 0.01 mV | - | ZA 9007-FS |
| Moisture sensor for materials | FH A696-MF | 0 … 50.0 | % | 0.1% | - | |
| Differential pressure | FD A612-SR | 0 … 1000 | mbar | 0.1 mbar | - | |
| Barometer | FD A612-SA | 0.0 … 1050 | mbar | 0.0 mbar | - | |
| Pressure transducers | FD-A602xx | (e.g.) 0.00 … 10.00 | bar | 0.01 bar | - | |
| Force transducer | FK Axxx | 0.0 … 50.00 | kN | 0.01 kN | - | |
| Displacement transducers | FW Axxx | 0.0 … 150.00 | mm | 0.01 mm | - | |
| Tachometer | FU A919-2 | 8 … 30000 | rpm | 1 rpm | - | ZA 9909-AK4U |
| Function values: |
| Difference | | | | | - | |
| Max value | | | | | - | |
| Minimum value | | | | | - | |
| Average value over time | | | | | - | |
| Average value over meas. pt. | | | | | - | |
| Sum over measuring points | | 0 … 65000 | | | - | |
| Total number of pulses | ZA 9909-AK2U | 0 … 65000 | | | - | |
| Pulses / print cycle | ZA 9909-AK2U | 0 … 65000 | | | - | |
| Alarm value | | 0.0 … 100.00 | % | | - | |
| Thermal coefficient | M(q)/M(ΔT) | | | | - | |
| Wet bulb globe temp. | (0.1TT+0.7HT+0.2GT) | | | | - | |
| Measuring value: |
| Cold junction temperature | | | °C | | | |
| Number of average values: |
| Volume flow | | 0 … 65000 | m3/h | m3/h | | |