Technical Specification
| Probe Specific Technical Data | |
|---|---|
| Operative Range | -20 to +80°C 5 to 98% RH |
| Mechanical Design Multi-sensor module: Plug connection: | Approx. 6 x 14 x 3mm (overall) Width approx. 7mm |
| FHAD 46-Cx. Common Technical Data | |
|---|---|
| Humidity Measuring range - Sensor - Accuracy - Hysteresis - Nominal temperature - Sensor operating pressure - Response time T63 | 0 to 98 % RH CMOSens® technology ±3 % RH in range 10 to 90 % RH ±5 % RH in range 5 to 98 % RH at nominal temperature typical ±1 % RH +23 °C ±5K Atmospheric pressure typical 8 seconds at +25°C, 1 m/s (without filter) |
| Temperature Sensor - Accuracy - Reproducibility - Response time T63- | CMOSens® technology typical ±0.2 K at +5 to 60°C maximum ±0.4 K at +5 to +60 °C maximum ±0.7 K at -20 to +80 °C typical ±0.1 K typical 20 seconds (without filter) |
| Digital atmospheric pressure sensor (integrated in multi-sensor module) Measuring range - Accuracy - | 700 to 1100 mbar ±2.5 mbar (at 23°C ±5K) |
| ALMEMO® D6 plug Refresh time - Supply voltage - Current consumption - | 1 second for all four channels 6 to 13 VDC 5 mA |