Technical Specification
| Input sockets: | Safety sockets CAT III, 20 A, Ø 4 mm |
| Galvanic isolation: | 6 kV |
| Sampling rate: | 1 kHz internal |
| Refresh rate: | 0.5s |
| AC signals U,I: | only sinusoidal signals, no signals with phase angle control |
| Response threshold U,I: | Signal U and I > 1 % of fin. val. |
| Operating range U, I, P: | DC … 250 Hz |
| Measuring range U, I, P: | see Types |
| Resolution: | see Types |
| Overload: | see Types |
| Internal resistance: | see Types |
| Accuracy: | ±0.1% of fin. val. ±2 Digit |
| Measuring range frequency: Resolution: | 20 to 250 Hz 0.01 Hz |
| Nominal conditions: | Alternating signal: sinusoidal 50 Hz, 23 °C ±2 K, 10 to 90% RH (non-condensing) |
| Temperature drift: | max. 0.003 %/K (30 ppm/K) |
| Suitable conditions: | +5 to +40 °C (Storage temperature: -20 to +60 °C), 10 to 90 % RH (non-condensing), maximum height above mean sea level: 2000 m |
| Housing: | ABS, 127 x 83 x 42 mm (LxWxH) |
| Connecting cable: | 2 m, permanently connected |
| ALMEMO® D6-plug: | for measuring channels, see Types, configuration via ALMEMO® device |
| Supply voltage: | 9 to 12 V via ALMEMO® device |
| Current consumption: | approx. 80 mA (plug and module) |



