Technical Specification
| Sensor | 2-beam infrared sensing element. Non-dispersive infrared technology (NDIR). |
| Measuring range | 0 to 5000 ppm Automatic compensation of pressure and temperature dependence of CO2 measurement with the built-in sensors. |
| Accuracy | ±(50 ppm +3 % of measured value) |
| Nominal conditions | 25°C, 1013 mbar |
| Switch-on time (initialization) | 15s |
| Response time t63 | 140s |
| Measuring interval | fixed 15 s as exponential moving average over 60 s (= 4 instantaneous values 15 s). |
| Range of application FYAD00CO2M3B05 FYAD00CO2M2B05 | -40 to 60°C -10 to 60°C 0 to 95% r.h. (non-condensing), 700 to 1100 mbar |
| Filter Cap | PTFE, diameter approx. 25 mm, length approx. 70 mm |
| Sensor tube FYAD00CO2M3B05 | stainless steel, diameter 12 mm, length approx. 130 mm |
| Sensor supply | via ALMEMO® D6 connector |
| Connection FYAD00CO2M3B05 FYAD00CO2M2B05 | permanently connected cable 2 m with ALMEMO® plug sensor mounted directly on ALMEMO® connector |
| Digital air pressure sensor (built-in) Measuring range Accuracy: | 700 to 1100 mbar typ. ± 2 mbar (at 25 °C) |
| Digital air temperature sensor (built-in) Measuring range Accuracy: typ. ALMEMO® D6 connector Measuring channels Refresh rate | -40 to +60°C typ. ± 0.5 °C (at 25 °C) Carbon dioxide concentration, air temperature, air pressure 15 sec. for all 3 channels |
| Supply voltage | 6 to 13 V DC |
| Current consumption | approx. 4 mA (avg), approx. 70 mA (max) |